Key result
Percutaneous closure of ventricular septal defects using the KONAR-MF multifunctional occluder achieved a 98.7% procedural success rate with a low incidence of major complications.
Why the study?
Transcatheter VSD closure offers an alternative to surgery, and the Lifetech Konar-Multifunctional Occluder has gained increasing application, motivating an evaluation of institutional experience with the device.
Is percutaneous VSD closure with the Konar-Multifunctional Occluder safe and effective in patients with muscular and perimembranous VSDs?
Observational (n=151)
No
Is percutaneous VSD closure with the Konar-Multifunctional Occluder safe and effective in patients with muscular and perimembranous VSDs?
Percutaneous VSD closure with the Konar-Multifunctional Occluder is highly successful (98.7%) and feasible, though associated with some valve regurgitation and vascular complications.
Supports MFO feasibility in pediatric VSD closure; extends single-center data but leaves open randomized comparisons to surgery.
Introduction Transcatheter device closure of ventricular septal defects (VSDs) offers an appealing and effective alternative to surgical repair. The Lifetech™ Konar-Multifunctional Occluder (MFO) VSD occluder has gained increasing application due to its versatility and promising outcomes. Objectives We aim to evaluate our experience with the MFO device for VSD closure. Methods We conducted a prospective analysis of clinical data from 151 patients who underwent percutaneous closure of muscular and perimembranous VSDs using the MFO device at our institution between November 2018 and September 2023. Comprehensive assessments of safety and procedural outcomes were performed. Results The patient's mean age was 55.4 ± 51.6 months (range, 6 months to 31 years), and the mean weight was 17.6 ± 11.9 kg (range, 5–86). Among the patients, 94 (62.3%) had perimembranous defects, while the remaining had muscular VSDs. The mean defect diameter was 4.8 ± 1.5 mm (range, 2–10). The retrograde approach was applied in 133 patients (88.7%). Device implantation was successful in 98.7% of patients. One procedure (0.7%) failed due to device migration, requiring surgical retrieval and VSD closure, and another patient with a significant residual shunt needed placement of an additional device in another session. The mean procedure time was 44 ± 2 min, with a mean fluoroscopy time of 12.8 ± 7.7 min. The mean follow-up duration was 11 ± 9.7 months (range, 6–35). Non-significant shunts were found in 32 patients (21.2%). Newly acquired valve regurgitation was observed in 16 patients (10.6%), including 11 patients (7.3%) with trivial-to-mild aortic regurgitation and 5 (3.3%) with moderate-to-severe tricuspid regurgitation. Electrophysiological adverse events occurred in 5 patients (3.3%), including nodal rhythm ( n = 3, 2%), intermittent heart block ( n = 1, 0.7%), and severe bradycardia ( n = 1, 0.7%). Vascular complications were documented in 13 patients (8.6%) including one developing chronic vascular occlusion. Conclusion Percutaneous VSD closure with the MFO device is a safe, effective, and feasible procedure via both antegrade and retrograde approaches.
No takes yet. Share an insight, caveat, or question.
Elafifi et al. (2025) conducted an observational in Ventricular septal defects (n=151). KONAR-MF multifunctional occluder was evaluated on Procedural success (proper device implantation without major complications or significant residual shunt). Percutaneous closure of ventricular septal defects using the KONAR-MF multifunctional occluder achieved a 98.7% procedural success rate with a low incidence of major complications.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: