Patent ductus arteriosus (PDA) with left-to-right shunting modifies the Doppler flow pattern of the intracranial circulation. The ability of increases in cerebral blood flow velocity (CBFV) to predict shunt resolution during PDA occlusion was evaluated. A 2 MHz transcranial Doppler (TCD) monitored diastolic and mean CBFV, plus the systolic/mean CBFV ratio in the middle cerebral artery from before (baseline) to immediately after PDA occlusion. Shunt resolution was verified by echocardiography and/or angiography. A minimum of 40% increase in diastolic-CBFV from baseline was considered successful resolution. Patients were age-stratified into group I (< 15 months; n = 23) and group II (> 15 months; n = 10). Thirty-three children were studied (age, 0.1 to 109 months) during surgical (n = 22) or coil occlusions (n = 11). Transcranial Doppler successfully identified shunt resolution in 78% of cases in group I, as compared to 0% in group II (p < 0.01). Identification rate decreased from 79% in cases of minimum ductal diameter of 3 mm (n = 19) to 21% in smaller ductuses (n = 14) (p < 0.01). Body weight and left-atrium size (p = 0.004) in group I and PDA diameter in group II (p = 0.02), were the only preoperative ductal parameters associated with diastolic-CBFV changes after ductus occlusion. Transcranial Doppler detects shunt resolution in infants with moderate to large PDAs.