Contrast-detail measurements were performed on a computed radiography imaging system as a function of detector entrance air kerma over the dose range from 0.743 microGy (0.085 mR) to 277 microGy (31.8 mR). A theoretical model of contrast-detail behaviour for a photostimulable phosphor computed radiography system has been derived, which is based on a modified version of the Rose theory of threshold detection. Included in the model are both system and x-ray quantum noise terms, as well as the response of the eye. The zero-frequency noise power of the computed film images was measured with a double-beam scanning microdensitometer. For a given detector dose, good agreement was found between the predicted and measured data when this measurement of system noise was included in the model. The contrast-detail results obtained for the computed radiography system were also compared with contrast-detail results for an image intensifier-TV based digital imaging system and a conventional film-screen system.