Key result
Complete removal or replacement of the VNS system, including the lead and coils, was feasible and safe in all 35 patients, with no incomplete removals.
Why the study?
Is complete removal or replacement of the VNS system safe and feasible in patients with refractory epilepsy?
Observational (n=35)
No
Is complete removal or replacement of the VNS system safe and feasible in patients with refractory epilepsy?
Complete removal or replacement of the VNS system is feasible and generally safe, though it carries a small risk of complications such as vocal cord paralysis and jugular vein laceration.
Supports VNS removal or replacement feasibility in refractory epilepsy; leaves open need for prospective validation.
BACKGROUND: With the growing use of vagus nerve stimulation (VNS) as a treatment for refractory epilepsy, there is a growing demand for complete removal or replacement of the VNS system. We evaluate the safety and efficacy of complete removal or replacement of the VNS system and provide an extensive description of our surgical technique. METHODS: We retrospectively reviewed our patient registry for all VNS surgeries performed between January 2007 (the year of our first complete removal) and May 2014. In order to assess patient satisfaction, a written questionnaire was sent to patients or their caregivers. Additionally, we reviewed all literature on this topic. RESULTS: The VNS system was completely removed in 22 patients and completely replaced in 13 patients. There were no incomplete removals. Revision surgery was complicated by a small laceration of the jugular vein in two patients and by vocal cord paralysis in one patient. Seizure frequency was unaltered or improved after revision surgery. Electrode-related side effects all improved after revision surgery. Twenty-one studies reported a total of 131 patients in whom the VNS system was completely removed. In 95 patients, the system was subsequently replaced. The most frequently reported side effect was vocal cord paresis, which occurred in four patients. CONCLUSIONS: Complete removal or replacement of the VNS system including lead and coils is feasible and safe. Although initial results seem promising, further research and longer follow-up are needed to assess whether lead replacement may affect VNS effectiveness.
No takes yet. Share an insight, caveat, or question.
Aalbers et al. (2015) conducted an observational in Refractory epilepsy (n=35). Complete removal or replacement of VNS system was evaluated on Complete removal of VNS system. Complete removal or replacement of the VNS system, including the lead and coils, was feasible and safe in all 35 patients, with no incomplete removals.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: