The Quantec Spinal Imaging System (QSIS) is a raster stereography used to measure three-dimensional trunk images. The Q angle, a coronal plane measurement generated by the Quantec Spinal Imaging System (QSIS), was compared with the Cobb angle in assessment of scoliosis curve magnitude. One hundred forty-nine patients with idiopathic scoliosis were evaluated using both the Quantec system and plane radiographs. The Cobb and Q angles demonstrated significant correlation in the thoracic region (r = 0.65, p < 0.05), lumbar region (r = 0.63, p < 0.05), and in the thoracolumbar region (r = 0.70, p < 0.05). The difference between the Q and Cobb angles was small when the Cobb angle was <21 degrees with less than 6 degrees of axial surface rotation, as measured by the QSIS method. For smaller curves with minimal rotation, there is close correlation between the Cobb angle and the Quantec angle.
No takes yet. Share an insight, caveat, or question.
Thometz et al. (2000) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: