Answers are grounded in this paper and cite their sources.
A liquid chromatography/tandem mass spectrometry (LC/MS/MS) method was developed and validated to determine the concentrations of adefovir [9-(2-phosphonylmethoxyethyl)adenine, PMEA] in human plasma. After one-step protein precipitation of plasma samples by methanol, adefovir was analyzed by LC/MS/MS using positive electrospray ionization. Chromatography was performed on a C18 column. The extraction recoveries of adefovir were found to be 85.1-89.3%. Adefovir was stable under routine laboratory conditions. A minimal matrix effect resulting in a slight ionization enhancement of adefovir (<10.9%) was observed, which did not markedly affect the behavior of the calibrations curves and accuracy and precision data. The method had a chromatographic run time of 7.8 min and a linear calibration curve over the concentration range 1.5-90 ng/mL for adefovir. The lower limit of quantification of the method was 1.5 ng/mL. The intra- and inter-day precision was less than 8.4%. These results indicated that this LC/MS/MS method has high selectivity and efficiency, and acceptable accuracy, precision and sensitivity. The validated LC/MS/MS method has been successfully used in a pharmacokinetic study in healthy volunteers treated with oral adefovir dipivoxil at 10 and 20 mg.