Key result
Short-term measurement of heart rate variability using a portable device in individuals with COPD demonstrated excellent intra- and interrater reliability, with intraclass correlation coefficients ranging from 0.915 to 1.000.
Why the study?
Heart rate variability assesses cardiac autonomic control in COPD but is evaluator-dependent and prone to methodological biases.
What is the inter- and intrarater reliability of short-term heart rate variability measurements in individuals with COPD?
Observational (n=51)
No
What is the inter- and intrarater reliability of short-term heart rate variability measurements in individuals with COPD?
Effect estimate: ICC 0.915-1.000
Short-term measurement of HRV using a portable device in COPD patients demonstrates acceptable intra- and interrater reliability, supporting its clinical and scientific use.
Individuals with COPD have an imbalance of cardiac autonomic control. In this context, HRV is considered an important tool for assessing cardiac sympathetic and parasympathetic balance, however it is a dependent evaluator measure and subject to methodological biases that may compromise the interpretation of results. This study examines the inter- and intrarater reliability of HRV parameters derived from short-term recordings in individuals with COPD. Fifty-one individuals of both genders with COPD clinical diagnosis confirmed by the pulmonary function test and aged ≥ 50 years were included. The RR interval (RRi) were recorded during a 10 min period on supine position using a portable heart rate monitor (Polar® H10 model). The data were transferred into Kubios® HRV Standard analysis software and analyzed within the stable sessions containing 256 sequential RRi. The intraclass correlation coefficient (ICC) ranged from 0.942 to 1.000 according to the intrarater analyse by Researcher 01 and 0.915 to 0.998 to the intrarater by Researcher 02. The interrater ICC ranged from 0.921 to 0.998. The coefficient of variation was up to 8.28 for Researcher 01 intrarater analysis, 9.06 for Researcher 02 intrarater analysis and 13.07 for interrater analysis. The measurement of HRV using a portable heart rate device in individuals with COPD present acceptable values of intra- and interrater reliability, supporting the use of HRV in the clinical and scientific scenario. Furthermore, it is important that the data analysis be performed by the same experienced evaluator.
No takes yet. Share an insight, caveat, or question.
Santos‐de‐Araújo et al. (2022) conducted an observational in Chronic Obstructive Pulmonary Disease (COPD) (n=51). Short-term measurement of heart rate variability was evaluated on Inter- and intrarater reliability of HRV parameters (Intraclass correlation coefficient) (ICC 0.915-1.000). Short-term measurement of heart rate variability using a portable device in individuals with COPD demonstrated excellent intra- and interrater reliability, with intraclass correlation coefficients ranging from 0.915 to 1.000.