PURPOSE: We studied whether a 10-MHz transvaginal transducer improves the diagnostic certainty that a small intrauterine fluid collection is a true gestational sac. METHODS: Over a 6-week period, women who presented with a positive pregnancy test and a fluid collection (devoid of any characteristics such as yolk sac or fetal pole) in the uterus seen with a standard 6-7-MHz transvaginal transducer were entered in the study. The patients were immediately rescanned with a 10-MHz transvaginal probe, and characteristics of the fluid collection using this probe were noted. Specifically, the visualization of a yolk sac or the intradecidual or double decidual sign was considered an objective improvement in the certainty that the fluid collection was a gestational sac. RESULTS: Twelve patients presented with a positive pregnancy test and a fluid collection in the uterus seen with a 6-7-MHz probe. Eight of these fluid collections were smaller than 1 cm in mean diameter, and all 8 of these patients had an objective improvement in the diagnosis of an intrauterine pregnancy using the 10-MHz probe. The other 4 patients had fluid collection 1-2 cm in mean diameter seen at 6-7 MHz. The 10-MHz probe improved the diagnostic confidence in 2 of these 4 patients. CONCLUSIONS: In patients with early pregnancies who have questionable gestational sacs on sonography with standard 6-7-MHz transducers, the 10-MHz probe improves the diagnostic confidence of the presence of an intrauterine gestational sac.