Key result
Left heart assistance using a vibrating flow pump at 30 Hz significantly increased brain blood flow compared with other drive frequencies at the same total cardiac output.
Why the study?
Does left heart assistance with a vibrating flow pump (VFP) at 30 Hz increase brain blood flow compared to other drive frequencies in healthy adult goats?
Does left heart assistance with a vibrating flow pump (VFP) at 30 Hz increase brain blood flow compared to other drive frequencies in healthy adult goats?
A vibrating flow pump VAD system operating at 30 Hz significantly increases brain blood flow in a goat model, suggesting potential for controlling end-organ perfusion.
In the clinical usage of the ventricular assist device (VAD), multiple organ failure becomes an important problem. To improve the clinical record of the VAD, another organ function may be vitally important. For that reason, we have been developing a VAD system aiming at improving another organ's function. Development of the vibrating flow pump (VFP), which can generate a very unique flow pattern from 10 Hz to 50 Hz, was ongoing in our Institute. In order to evaluate brain blood flow and oxygen consumption, HbO2 was measured with a NIRO monitoring device in healthy adult goats. Four goats were anesthetized with halothane inhalation; then left thoracotomy was performed for the left heart bypass. HbO2 of the brain was measured by recording of the hemodynamic variables during left heart assistance with the VFP system. During left heart bypass with the VFP system, hemodynamic parameters stayed within normal range, and satisfactory pump output was easily obtained. Pump output stayed within 20-40% bypass to evaluate the effect of high frequency oscillated assist flow on brain blood flow during the same cardiac output. Interesting results were observed during the experiments. During 30 Hz drive of the VFP left heart assistance, HbO2 suggested that brain blood flow significantly increased compared with another drive frequency assistance during the same total cardiac output. These results suggest that we can control the brain blood flow with a totally implantable VAD system such as the VFP system.
No takes yet. Share an insight, caveat, or question.
Yambe et al. (2000) studied Healthy (n=4). Vibrating flow pump (VFP) system vs. Other drive frequencies was evaluated on Brain blood flow (measured by HbO2). Left heart assistance using a vibrating flow pump at 30 Hz significantly increased brain blood flow compared with other drive frequencies at the same total cardiac output.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: