Key result
The Doppler wave index (EId) demonstrated a strong correlation with the blood flow volume index (EIV) in a phantom model (r = 0.97, p<0.01).
Why the study?
Does the Doppler waveform area index correlate with blood flow volume in a phantom model?
Does the Doppler waveform area index correlate with blood flow volume in a phantom model?
Answers are grounded in this paper and cite their sources.
The Doppler waveform area index strongly correlates with blood flow volume in a phantom model, suggesting potential utility for assessing valve efficiency.
Objective: To demonstrate a significant correlation between the Doppler waveform and blood flow volume in the assessment of valve efficiency. In particular the objective was to ratify use of the ‘area index’ of the wave form. Design: Pearson's correlation and zero line assessment were used. Using a phantom, multiple tests were performed so that a wide spectrum of flow volumes could be compared with corresponding Doppler waves. Setting: Private diagnostic imaging clinic with vascular surgical association. Main outcome measures: The hypothesis was that there should be a significant correlation between the Doppler wave and the blood flow volume and, in particular, the Doppler area. Results: The Doppler wave index EI d demonstrated the highest correlation coefficient ( r = 0.97, p<0.01) when compared with the volume index EI V . Conclusion: While there is a strong correlation between EI d and flow volume, additional study is needed to determine its place in the clinical setting.
No takes yet. Share an insight, caveat, or question.
Beckwith et al. (1993) studied Valve efficiency assessment. Ultrasonic Doppler wave forms vs. Blood flow volume was evaluated on Correlation between the Doppler wave and blood flow volume (r = 0.97, p=<0.01). The Doppler wave index (EId) demonstrated a strong correlation with the blood flow volume index (EIV) in a phantom model (r = 0.97, p<0.01).
Synapse has enriched 10 related papers. Consider them for comparative context: