A novel atomic fluorescence spectrometer is reported for the determination of hydride-forming elements, principally arsenic and selenium. A miniature argon–hydrogen diffusion flame was used as the atomizer and the analyte elements were introduced as their gaseous hydrides from a fully automated continuous hydride generator. The hydrogen for the flame was chemically generated as a by-product of the sodium tetrahydroborate reduction. Excitation was achieved using a boosted-discharge hollow cathode lamp. Fluorescence wavelengths of interest were selected using an interference filter. A solar blind photomultiplier was used as the detector. The instrument is notably compact and capable of full automation by connection to a personal computer using a DIO card. Optimization of the instrumentation is described. Detection limits (3σ) of 0.10 and 0.05 µg l–1 for arsenic and selenium, respectively, are reported, as are a number of analyses of certified reference water samples, which confirm the excellent accuracy and precision of the new instrumentation.
No takes yet. Share an insight, caveat, or question.
Corns et al. (1993) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: