Key result
Combined fluoroscopy and intracardiac echocardiography enabled real-time anatomic localization and successful percutaneous retrieval of a dislodged, almost radiolucent catheter fragment.
Case Report
Combined fluoroscopy and intracardiac echocardiography can serve as a crucial aid in the percutaneous retrieval of intracardiac foreign bodies that are difficult to visualize and adjacent to cardiac valves.
Answers are grounded in this paper and cite their sources.
May aid retrieval of radiolucent intracardiac fragments; single case leaves open need for validation.
This report details a method of percutaneous, transluminal retrieval of an intracardiac foreign body using fluoroscopy in combination with intracardiac echocardiography. During retrieval, intracardiac echocardiography (ICE) provided real-time anatomic localization of a constantly moving, almost radiolucent micropuncture coaxial dilator fragment with respect to the tricuspid and pulmonary valves. This method may serve as a crucial aid in retrieval of intracardiac foreign bodies that are difficult to see with fluoroscopy and which may be adjacent to cardiac valves.
No takes yet. Share an insight, caveat, or question.
Mitsopoulos et al. (2015) conducted a case report in Intracardiac foreign body (dislodged catheter). Combined fluoroscopy and intracardiac echocardiography (ICE) was evaluated on Retrieval of an intracardiac foreign body. Combined fluoroscopy and intracardiac echocardiography enabled real-time anatomic localization and successful percutaneous retrieval of a dislodged, almost radiolucent catheter fragment.