Answers are grounded in this paper and cite their sources.
A comparative method for the quantitative analysis of the ratio of oxonium fragment (reporter) ions derived from sialic acid and N-acetylhexosamine residues on a large intact glycoprotein, the B domain of recombinant human factor VIII (rhFVIII), was developed. The method utilized liquid chromatography-electrospray ionization mass spectrometry (LC-ESI MS) on a single-quadrupole instrument. During development, systematic approaches such as full-matrix and simplex strategies were used for the optimization of the signal-to-noise ratio by controlling source temperature and cone voltage. The method was found to be precise (RSD = 0.84%), sensitive (capable of differentiating 1 sialic acid residue change among at least 29 sialic acids on a 103-kDa glycoprotein that is 38% carbohydrate), applicable to a wide range of loading (11.6-372 micrograms of FVIII), and accurate according to a comparison to matrix-assisted laser desorption-ionization time-of-flight mass spectrometry. Combining the method with enzymatic removal of N-glycans, selective O-glycan analysis was also performed leading to differential fragment ion analysis ascribed to N- and O-glycans. Quantitative ESI in-source dissociation MS combined with LC can generally be used for glycoproteins, as one of the indicators, to compare the distribution of carbohydrate residues over N- and O-glycans, to investigate their isoforms, and compare batch-to-batch characteristics of biopharmaceuticals.
No takes yet. Share an insight, caveat, or question.
Mazsaroff et al. (1997) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: