A novel cavity resonator method is proposed for measuring the complex permittivity of a dielectric plate in the millimeter-wave region. In this method, the dielectric plate is loaded at the end of the cavity resonator. The relative permittivityε'and loss tangenttanδare determined from measured resonant frequency and unloadedQof theTE₀₁₁mode. This structure prevents excessive lowering of the resonant frequency due to the placement of the dielectric plate sample in the cavity. The suppression of spurious modes and the design of the cavity resonator are discussed. The values ofε'andtanδwere measured in the millimeter-wave range for low-temperature co-fired ceramic (LTCC), Al₂O₃, and sapphire plates. The temperature dependence of the complex permittivity of the LTCC plate was measured at 34 GHz. The measurement accuracies are estimated to be within 0.6% forε', and within 10% fortan δof 10⁻³.
No takes yet. Share an insight, caveat, or question.
Yoshikawa et al. (2008) studied this question.
Synapse has enriched 2 closely related papers on similar clinical questions. Consider them for comparative context: