PURPOSE: To compare the short-term outcome of the Multigen Plus system using a symmetric ceramic femoral component with that of 2 metallic total knee replacement (TKR) systems. METHODS: 60 patients (62 knees) with osteoarthritis or rheumatoid arthritis underwent primary TKR using the Genia system (n=20), the Multigen Plus system (n=17), and the Multigen Plus system with a ceramic femoral component (n=25). Outcome was evaluated pre- and post-operatively (at 3, 12, and 24 months) using the Hospital for Special Surgery (HSS) score, the Western Ontario and McMaster Universities (WOMAC) score, and the Short-Form 36 (SF-36) score. RESULTS: The 3 groups did not differ significantly at all 4 time points in terms of the HSS score, WOMAC score, and SF-36 score. The range of motion was significantly less in patients treated with the Genia system. Postoperatively, the HSS and WOMAC scores improved significantly in all 3 groups, whereas the SF-36 score improved significantly (at 12 and 24 months) only in patients treated with the Multigen Plus system with a ceramic femoral component. The implant position in most patients was optimal. All radiolucent lines were within 1 mm and did not progress. CONCLUSION: The short-term outcome of the 3 TKR systems was comparable. Ceramic femoral components performed comparably to traditional metallic designs.
No takes yet. Share an insight, caveat, or question.
Bergschmidt et al. (2013) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: