Key result
Contact and video photoplethysmography waveforms exhibited primary features of jugular vein waveforms during physiological testing in 2 volunteers, demonstrating potential for non-invasive monitoring.
Why the study?
Central venous pressure assesses hemodynamic status but requires invasive measurement, making non-invasive alternatives of great clinical utility.
Can contact and contactless photoplethysmography (PPG) characterize venous system fluid motion in the internal jugular vein during physiological testing?
Case Report (n=2)
Can contact and contactless photoplethysmography (PPG) characterize venous system fluid motion in the internal jugular vein during physiological testing?
Contact and video photoplethysmography show potential for non-invasive monitoring of venous blood flow and central venous pressure estimation.
Hypothesis-generating for non-invasive jugular venous monitoring with PPG; leaves open validation in clinical cohorts.
Central venous pressure is an estimate of right atrial pressure and is often used to assess hemodynamic status. However, since it is measured invasively, non-invasive alternatives would be of great utility. The aim of this preliminary study was a) to investigate whether photoplethysmography (PPG) can be used to characterize venous system fluid motion and b) to find the model for venous blood volume modulations. For this purpose, we monitored the internal jugular veins using contact (cPPG) and video PPG during clinically validated physiological tests: abdominojugular test (AJT) and breath holding (BH). Video PPG and cPPG signals were captured simultaneously on the left and right sides of the neck, respectively. ECG was also captured using the same clinical monitor as cPPG. Two volunteers underwent AJT and BH with head up/down, each with: baseline (15s), experiment (15s), and recovery (15s). Video PPG was split into remote PPG (rPPG) and micromotion detection. All signal modalities were significantly affected by physiological testing. Moreover, cPPG and micromotion waveforms exhibited primary features of jugular vein waveforms and, therefore, have great potential for venous blood flow monitoring. Specifically, remote patient monitoring applications may be enabled by this methodology, facilitating physical collection without a specially trained care provider.
No takes yet. Share an insight, caveat, or question.
Saiko et al. (2024) reported a case report. Contact and video photoplethysmography (PPG) was evaluated on Venous system fluid motion and blood volume modulations. Contact and video photoplethysmography waveforms exhibited primary features of jugular vein waveforms during physiological testing in 2 volunteers, demonstrating potential for non-invasive monitoring.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: