Key result
An HPLC-DAD analytical method for determining anthracycline drug residuals on stainless-steel equipment demonstrated a recovery of 90.1% with an LOD of 0.047 μg/mL and LOQ of 0.155 μg/mL.
The developed HPLC-DAD method is valid for determining anthracycline drug residuals on stainless-steel equipment for cleaning validation.
May support cleaning validation in anthracycline handling; leaves open broader clinical adoption without prospective confirmation.
Standard cleaning procedures of production line equipment were verified after manufacture of cytostatic injections containing Anthracycline derivate substance. Residual content of Anthracycline drug substance on stainless-steel equipment surface was determined using swab sampling with a specific HPLC-DAD analysis. The acceptance limit was decided as 200.0 μg/100 cm2. Recovery from the stainless-steel surface was 90.1%. Linearity of the method was observed in the concentration range of 0.155–194 μg/mL when estimated using Zorbax TMS (5 μm, 0.25 m × 4.6 mm ID) column at 1.3 mL/min flow rate and 254 nm (DAD 190–600 nm). The mobile phase consisted of lauryl hydrogen sulphate solution (3.7 g/L) : methanol : acetonitrile (54 : 16 : 30, v/v/v) with pH adjusted to 2.5 using phosphoric acid (85%). The LOD and LOQ for Anthracycline derivate were found to be 0.047 and 0.155 μg/mL, respectively. The method validation confirmed the method provides acceptable degree of selectivity, linearity, accuracy, and precision for the intended purposes.
No takes yet. Share an insight, caveat, or question.
Slivová et al. (2015) studied this question. HPLC-DAD analysis for Anthracycline drug residual was evaluated on Recovery from stainless-steel surface. An HPLC-DAD analytical method for determining anthracycline drug residuals on stainless-steel equipment demonstrated a recovery of 90.1% with an LOD of 0.047 μg/mL and LOQ of 0.155 μg/mL.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: