Mercury ion (Hg 2+ ) is extremely toxic even at very low concentrations and its detection mainly relies on bulky and high-cost analytical instruments. Here, we introduce a fast and ultrasensitive biosensing method developed by the integration of fiber optic particle plasmon resonance and a highly selective molecular beacon with a stem-loop DNA structure immobilized on the unclad surface of an optical fiber. In the presence of Hg 2+ ions and a free assisting DNA probe, the stem-loop opens and forms thymine–Hg 2+ –thymine complexes. A DNA reporting probe conjugated to gold nanoparticles is used as a label to interrogate the recognition event by binding with the terminal DNA binding domain of the opened stem-loop. The method provides a wide linear range of at least 5 orders from 10 –14 to 10 –9 M and an extremely low detection limit of 4.37 fM (0.876 ppq). It takes less than 15 min to analyze the concentration of Hg 2+ ions in aqueous samples. The method is desirable for point-of-use applications because of its low-cost instrumentation and sensor chips, easy operation, and portability.
No takes yet. Share an insight, caveat, or question.
Fan et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: