Answers are grounded in this paper and cite their sources.
X-ray spectrometers used in high energy density plasma experiments must provide high time, space, and spectral resolution while overcoming the difficulties imposed by x-ray background, debris, and mechanical shocks. At the Z facility these problems are addressed using a suite of elliptical crystal spectrometers. The elliptical geometry isolates the detector from the line of sight with a slit placed at the elliptical focus, while the sensitivity enables locating the crystal 2–4 m from the plasma source. Space and time resolution are obtained by using an array of slits to project one dimensional plasma images onto the crystal and recording the spectrally dispersed images with a gated microchannel plate detector.
No takes yet. Share an insight, caveat, or question.
Lake et al. (2004) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: