Key result
A machine learning model using noninvasive pulsatile cranial expansion waveforms estimated intracranial pressure with a mean absolute error of 3.00 mmHg compared to invasive monitoring.
Why the study?
Noninvasive methods for intracranial pressure monitoring have emerged, but none has successfully replaced invasive techniques.
Does a machine learning model using waveforms from a cranial extensometer device accurately estimate intracranial pressure compared to invasive monitoring in neurocritical patients with acute brain injuries?
Observational (n=112)
Yes
Does a machine learning model using waveforms from a cranial extensometer device accurately estimate intracranial pressure compared to invasive monitoring in neurocritical patients with acute brain injuries?
A machine learning model using noninvasive cranial expansion waveforms can estimate intracranial pressure with a mean absolute error of 3.00 mmHg compared to invasive monitoring.
May support noninvasive ICP estimation in neurocritical care; hypothesis-generating and requires prospective validation before adoption.
Noninvasive methods for intracranial pressure (ICP) monitoring have emerged, but none has successfully replaced invasive techniques. This observational study developed and tested a machine learning (ML) model to estimate ICP using waveforms from a cranial extensometer device (brain4care [B4C] System). The model explored multiple waveform parameters to optimize mean ICP estimation. Data from 112 neurocritical patients with acute brain injuries were used, with 92 patients randomly assigned to training and testing, and 20 reserved for independent validation. The ML model achieved a mean absolute error of 3.00 mmHg, with a 95% confidence interval within ±7.5 mmHg. Approximately 72% of estimates from the validation sample were within 0-4 mmHg of invasive ICP values. This proof-of-concept study demonstrates that noninvasive ICP estimation via the B4C System and ML is feasible. Prospective studies are needed to validate the model's clinical utility across diverse settings.
No takes yet. Share an insight, caveat, or question.
Frigieri et al. (2025) conducted an observational in Acute brain injury (n=112). Machine learning model using brain4care (B4C) System waveforms vs. Invasive ICP monitoring was evaluated on Mean absolute error (MAE) of estimated ICP (MAE 3.00 mmHg, 95% CI ±7.5). A machine learning model using noninvasive pulsatile cranial expansion waveforms estimated intracranial pressure with a mean absolute error of 3.00 mmHg compared to invasive monitoring.