Scanning near-field optical microscopes (SNOM) using the tetrahedral-tip (T-tip) with scanning tunnelling microscopy (STM) distance control have been realized in transmission and reflection mode. Both set-ups used ordinary STM current-to-voltage converters allowing measurement of metallic samples. In the transmission mode, a resolution of 10 nm to 1 nm with regard to material contrast can be achieved on binary metal samples. Because of the great near-field optical potential of the T-tip with respect to the optical resolution, it is a challenging task to find out whether these results can be transferred to non-metallic sample systems as well. This paper reports on a newly designed SNOM/STM transmission mode set-up using the tetrahedral-tip. It implements a sensitive current-to-voltage converter to widen the field of measurable sample systems. Beyond this, mechanical and optical measuring conditions are substantially improved compared to previous set-ups. The new set-up provides a basis for the routine investigation of metal nanostructures and adsorbed organic monolayers at resolutions in the 10 nm range.