Answers are grounded in this paper and cite their sources.
The applicability of the subtractive anodic stripping voltammetry (SASV) using the square‐wave mode at the silver‐gold alloy electrode has been studied for thallium determination in the presence of large amount of lead and cadmium in natural samples. 10 mmol L−1 perchloric acid was found as the most suitable supporting electrolyte for determination in synthetic solutions. The thallium peak was separated about 200 mV from Cd+Pb peak. Diethylenetriaminepentaacetic acid addition was necessary to determine thallium at the silver‐gold alloy electrode in digested plant and sediment. The determination limit was equal to 1.4 μg L−1. The method was validated by the inter‐method comparison (ICP‐MS).