Experimental study demonstrates sub-quantum-limited noise in a cascaded parametric amplifier, highlighting a path toward direct quantum readout without background subtraction.
Phase-sensitive parametric devices facilitate quadrature-selective amplification with the possibility for sub-quantum-limited noise performance. In this work, we investigate the operation of a SQUID-based Josephson parametric amplifier (JPA), comparing its gain and noise performance in the phase-preserving and phase-sensitive regimes. The device, fabricated using VTT SWAPS technology, is driven in a three-wave mixing configuration and characterized in a reflection-based measurement setup at millikelvin temperatures. To directly probe the noise performance at low JPA gains, we employ a cascaded amplification scheme in which a traveling-wave parametric amplifier (TWPA) provides low-noise preamplification of the JPA output. In a phase-preserving operation, the JPA exhibits near-quantum-limited performance with a system noise temperature of 351± 53 351 ± 53 mK at 6 GHz. In contrast, phase-sensitive operation yields a minimum system noise temperature of 94± 12 94 ± 12 mK, well below the standard quantum limit of 288 mK. Our results demonstrate that a JPA–TWPA amplifier cascade opens the door to direct, high-fidelity probing of quantum devices without the need for background noise subtraction.
No takes yet. Share an insight, caveat, or question.
Lilja et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: