An assay for the measurement of human growth hormone in urine (u-HGH) has been developed using commercially available reagents and its analytical variables have been investigated. The procedure is sensitive to 0.50 muU/L urine, has an interassay precision of 6.2-7.5 (CV%) over the standard curve, and shows good parallelism and recovery. Urine samples for the assay can be collected in plain plastic bottles, stored at room temperature for up to 24 h and, after adding preservative, at -20 C before assay for three months. There was a good correlation between the responses of HGH in serum and in urine during clonidine and insulin stimulation tests.
No takes yet. Share an insight, caveat, or question.
Evans et al. (1989) studied this question.