Key result
During a 20-minute LAD occlusion in swine, a multiparameter fiber optic sensor detected a decline in myocardial pH (7.36 to 6.85) and PO2 (78 to 6 torr), and a rise in PCO2 (57 to 154 torr).
Why the study?
Does a multiparameter fiber optic sensor reliably detect changes in intramyocardial perfusion during coronary occlusion in a swine model?
Does a multiparameter fiber optic sensor reliably detect changes in intramyocardial perfusion during coronary occlusion in a swine model?
A multiparameter fiber optic sensor can reliably detect ischemic changes in myocardial pH, PCO2, and PO2 during coronary occlusion in a swine model, suggesting potential utility for guiding myocardial protection during cardiac surgery.
Answers are grounded in this paper and cite their sources.
Validates multiparameter detection during porcine ischemia; leaves open translation to cardiac surgical monitoring.
OBJECTIVES: The objective of this study was to characterize a multiparameter fiber optic sensor for detection of changes in intramyocardial perfusion and to demonstrate a method of determining critical values for pH, PCO2, and PO2 to indicate onset of anaerobic metabolism. METHODS: Six swine underwent a 20-minute occlusion of the left anterior descending coronary artery (LAD). Myocardial pH, PCO2, and PO2 were measured continuously in the LAD and left circumflex coronary artery (CFX) territories. Critical values for each parameter were calculated from these data. RESULTS: During occlusion LAD myocardial pH declined from 7.36 +/- 0.04 to 6.85 +/- 0.04; PCO2 rose from 57.0 +/- 2.9 to 154.0 +/- 18.0 torr, PO2 fell from 78 +/- 20 to 6 +/- 5 torr. No myocardial pH or PCO2 changes were observed in the CFX region, however, CFX PO2 was affected in some animals during LAD occlusion and release. Methods for determining the ischemic threshold from these sensor data are presented. CONCLUSIONS: Multiparameter fiber optic sensors reliably respond to coronary occlusion and thus have the potential to help guide myocardial protection strategies for both on- and off-pump cardiac surgery.
No takes yet. Share an insight, caveat, or question.
Soller et al. (2004) studied Myocardial ischemia (n=6). Multiparameter fiber optic sensor monitoring vs. Non-occluded left circumflex coronary artery (CFX) territory was evaluated on Changes in myocardial pH, PCO2, and PO2. During a 20-minute LAD occlusion in swine, a multiparameter fiber optic sensor detected a decline in myocardial pH (7.36 to 6.85) and PO2 (78 to 6 torr), and a rise in PCO2 (57 to 154 torr).
Synapse has enriched 10 related papers. Consider them for comparative context: