BACKGROUND: The purpose of this study is to assess the long-term follow-up of cementless total knee arthroplasty with the trabecular metal (TM) monoblock tibial component at an average 10-year follow-up. This report is an extension of our previously reported series of 108 TM tibias reported in 2011 (Unger and Duggan, 2011). METHODS: Fifty-eight of the original 108 knees were available for review. Each follow-up patient was evaluated by radiologic and clinical Knee Society Scores. The average follow-up was 10.2 years. RESULTS: Our results indicate excellent long-term survivorship (96.5%) with 2 confirmed tibia revisions, and 1 femoral revision for periprosthetic fracture and 1 patella open reduction internal fixation. X-ray evaluation demonstrated one patient with 1 mm medial polyethylene wear and a nonprogressive 1 mm of radiolucency on the medial side. All the other tibial components showed full bone apposition and incorporation. Knee Society Scores were excellent in all the patients seen on follow-up. CONCLUSIONS: Long-term follow-up of TM monoblock tibia components confirm excellent survivorship and biologic implant fixation, with excellent outcomes and knee scores.