Does the Abbott HTLV III EIA accurately detect antibodies to HTLV III in patients with AIDS and blood donors?
Does the Abbott HTLV III EIA accurately detect antibodies to HTLV III in patients with AIDS and blood donors?
The Abbott HTLV III EIA is a highly sensitive and specific test for detecting HTLV III antibodies in donor blood and AIDS patients.
Supports EIA for HTLV-III detection in AIDS and donors; leaves open prospective validation before routine use.
An enzyme immunoassay (EIA), designed to detect antibodies to human T-cell lymphotropic virus type III (HTLV III), was evaluated. The antibody test was found to be highly sensitive; serum from 221 of 223 (99.1%) patients with acquired immune deficiency syndrome (AIDS) was positive for antibodies to HTLV III. In addition, the antibody test was found to be highly specific; approximately 99.75% of 20,720 serum or plasma samples from blood donors were negative for antibody to HTLV III. In most cases, the Western Blot analysis agreed well with the EIA. Eighty-one of 82 (98.8%) EIA-positive samples from patients with AIDS were Western Blot positive. Of the EIA-positive blood donors, 21 of 36 (58%) were detected and confirmed by Western Blot analysis. A solid-phase competitive EIA also was evaluated as an alternate procedure. The preliminary results indicate that this immunoassay has a high degree of sensitivity and specificity and could serve as an alternate procedure to detect antibody to HTLV III.