Key result
Heart rate in pacemaker-dependent patients correlated well with minute ventilation (correlation coefficient 0.72-0.95, P<0.0001), demonstrating its utility as a metabolic sensor.
Why the study?
Does a minute ventilation sensing rate responsive pacemaker accurately correlate heart rate to minute ventilation during exercise in patients with bradycardia?
Does a minute ventilation sensing rate responsive pacemaker accurately correlate heart rate to minute ventilation during exercise in patients with bradycardia?
Minute ventilation serves as a reliable metabolic sensor for rate-responsive pacing in patients with bradycardia.
Supports investigation of minute ventilation sensors for rate-responsive pacing; leaves open routine use pending larger prospective trials.
A minute ventilation sensing rate responsive pacemaker was implanted in 15 patients (8 males and 7 females) with bradycardia. The mean age was 72.8 +/- 8.7 years. The single chamber system measures transthoracic impedance between the tip electrode of a standard bipolar lead and the pulse generator case. In the adaptive mode the pulse generator calculates a rate responsive factor or slope during maximal exercise but functions as in the VVI mode. The patients exercised maximally on an upright cycle ergometer with the pacemaker programmed to VVI mode, adaptive mode, and rate responsive mode. Exercise and gas exchange data were collected continuously and analyzed using an automated breath-by-breath system. The slope, heart rate, and ventilation were measured every 20 seconds. Heart rate in pacemaker dependent patients correlated well to minute ventilation (correlation coefficient ranging from 0.72-0.95, P < 0.0001). This study demonstrates that minute ventilation is a good metabolic sensor in rate responsive pacing.
No takes yet. Share an insight, caveat, or question.
Abrahamsen et al. (1993) studied bradycardia (n=15). Minute ventilation sensing rate responsive pacemaker vs. VVI mode was evaluated on Correlation between heart rate and minute ventilation (correlation coefficient 0.72-0.95, p=< 0.0001). Heart rate in pacemaker-dependent patients correlated well with minute ventilation (correlation coefficient 0.72-0.95, P<0.0001), demonstrating its utility as a metabolic sensor.