BACKGROUND/OBJECTIVES: Heated intraperitoneal chemotherapy (HIPEC) kills cancer cells via thermal injury and improved chemotherapeutic cytotoxicity. We hypothesize that higher HIPEC flow rates improve peritoneal heating and HIPEC efficacy. METHODS: (1) A HIPEC-model (30.8 L cooler with attached extracorporeal pump) was filled with 37°C water containing a suspended 1 L saline bag (SB) wrapped in a cooling sleeve, creating a constant heat sink. (2) HIPECs were performed in a swine model. Inflow, outflow, and peritoneal temperatures were monitored as flow rates varied. (3) Flow rates and temperatures during 20 HIPECs were reviewed. RESULTS: Higher flow rates decreased time required to increase water bath (WB) and SB temperature to 43°C. With a constant heat sink, the minimum flow rate required to reach 43°C in the WB was 1.75 L/min. Higher flow rates lead to greater temperature gradients between the WB and SB. In the swine model, the minimum flow rate required to reach 43°C outflow was 2.5-3.0 L/min. Higher flows led to more rapid heating of the peritoneum and greater peritoneal/outflow temperature gradients. Increased flow during clinical HIPEC suggested improved peritoneal heating with lower average visceral temperatures. CONCLUSIONS: There is a minimum flow rate required to reach goal temperature during HIPEC. Flow rate is an important variable in achieving and maintaining goal temperatures during HIPEC.