We demonstrate a high-brightness compact 9 keV electron-impact microfocus x-ray source based on a liquid-gallium-jet anode. A approximately 30 W, 50 kV electron gun is focused onto the approximately 20 ms, 30 mum diameter liquid-gallium-jet anode to produce an approximately 10 microm full width at half maximum x-ray spot. The peak spectral brightness is >2 x 10(10) photons(s mm(2) mrad(2)x0.1% BW). Calculation and experiments show potential for increasing this brightness by approximately three orders of magnitude, making the source suitable for laboratory-scale x-ray crystallography and hard x-ray microscopy.