Answers are grounded in this paper and cite their sources.
An alternative spectroscopic approach for monitoring the temperature of aqueous solutions is presented. The method is based upon the temperature-induced spectral changes undergone by the second overtone (around 960 nm) of the near-infrared (NIR) water absorption band. Single and multilinear regression analysis are tested in order to evaluate the predictive ability of temperature. A linear dependence is found when measurements are performed at a single wavelength, but a lower prediction error is obtained when multilinear models are applied. No matrix effects produced by moderately concentrated common dissolved ions are found in a broad range of pH. A signal-to-noise ratio allows a precision of 0.5 degrees C for temperature monitoring. A prediction error of 0.77 degrees C (single linear regression) and 0.25 degrees C (multilinear approach) are achieved in a range from 15 to 90 degrees C. Advantages in terms of instrumentation and data analysis required are discussed.
No takes yet. Share an insight, caveat, or question.
Otal et al. (2003) studied this question.