A method has been developed to optimize the selection of beam weights for external beam radiation therapy treatment planning. This method uses a matrix transformation method to achieve a least-squares fit to doses at preselected points in a target volume. This method is more general than the conventional least-squares method in that it allows one to incorporate as constraints maximum allowable doses delivered to critical organs, as well as the requirement that all beam weights be non-negative. A computer algorithm DOSEOPT has been written which takes a matrix of doses to various points from beams whose field sizes and orientations have been predetermined, and performs the constrained least-squares fit to desired doses in the target volume and maximum doses to critical organs. Implementation of this algorithm and its application to several cases are discussed.
No takes yet. Share an insight, caveat, or question.
George Starkschall (1984) studied this question.