Key result
The in-silico multiphysics testbed accurately predicted a 23°C temperature increase at a resonant metal conductor tip, closely matching the 22°C experimental measurement.
Why the study?
The presence of implants limits MRI-based health monitoring in older adults, while the prevalence of implantable devices and conditions requiring MRI is increasing.
A novel multiphysics in-silico testbed has been developed and FDA-approved for evaluating MR-induced heating of orthopaedic implants.
Answers are grounded in this paper and cite their sources.
May facilitate safer MRI in patients with implants; leaves open prospective clinical validation of the testbed.
One limitation on the ability to monitor health in older adults using Magnetic Resonance (MR) imaging is the presence of implants, where the prevalence of implantable devices (orthopedic, cardiac, neuromodulation) increases in the population, as does the pervasiveness of conditions requiring MRI studies for diagnosis (musculoskeletal diseases, infections, or cancer). The present study describes a novel multiphysics implant modeling testbed using the following approaches with two examples:an in-silico human model based on the widely available Visible Human Project (VHP) cryo-section dataset;a finite element method (FEM) modeling software workbench from Ansys (Electronics Desktop/Mechanical) to model MR radio frequency (RF) coils and the temperature rise modeling in heterogeneous media.The in-silico VHP Female model (250 parts with an additional 40 components specifically characterizing embedded implants and resultant surrounding tissues) corresponds to a 60-year-old female with a body mass index (BMI) of 36. The testbed includes the FEM-compatible in-silico human model, an implant embedding procedure, a generic parameterizable MRI RF birdcage two-port coil model, a workflow for computing heat sources on the implant surface and in adjacent tissues, and a thermal FEM solver directly linked to the MR coil simulator to determine implant heating based on an MR imaging study protocol. The primary target is MR labeling of large orthopaedic implants. The testbed has very recently been approved by the US Food and Drug Administration (FDA) as a medical device development tool (MDDT) for 1.5 T orthopaedic implant examinations.
No takes yet. Share an insight, caveat, or question.
Noetscher et al. (2023) studied Orthopaedic implants requiring MRI (n=1). In-silico multiphysics implant modeling testbed vs. Experimental measurements was evaluated on Temperature rise prediction near implants during MRI scan. The in-silico multiphysics testbed accurately predicted a 23°C temperature increase at a resonant metal conductor tip, closely matching the 22°C experimental measurement.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: