Answers are grounded in this paper and cite their sources.
A sensitive and reliable method of liquid chromatography-electrospray ionization/tandem mass spectrometry (LC-ESI/MS/ MS) was developed and validated for determining 1,3-dimethylamylamine (1,3-DMAA) and 1,4-dimethylamylamine (1,4-DMAA) in geranium plants (Pelargonium graveolens). The sample was extracted with 0.5 M HCl and purified by liquid-liquid partition with hexane. The parameters for reverse-phase (C18) LC and positive ESI/MS/MS were optimized. The matrix effect, specificity, linearity, precision, accuracy and reproducibility of the method were determined and evaluated. The method was linear over a range of 0.10-10.00 ng/mL examined, with R(2) of 0.99 for both 1,3-DMAA and 1,4-DMAA. The recoveries from spiked concentrations between 5.00-40.00 ng/g were 85.1%-104.9% for 1,3-DMAA, with relative standard deviation (RSD) of 2.9%-11.0%, and 82.9%-101.8% for 1,4-DMAA, with RSD of 3.2%-11.7%. The instrument detection limit was 1-2 pg for both DMAAs. The quantification limit was estimated to be 1-2 ng/g for the plant sample. This method was successfully applied to the quantitative determination of 1,3- and 1,4-DMAA in both geranium plant and geranium oil.
No takes yet. Share an insight, caveat, or question.
Li et al. (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: