Answers are grounded in this paper and cite their sources.
This work demonstrates an electrostatically actuated resonant microcantilever fabricated in a complementary metal oxide semiconductor process and functionalized with a chemically sorbent polymer layer for the detection of volatile organic compounds. Deposition of the chemically sorbent layer is controlled through evaporation-driven assembly. Analytical and finite element analysis models of the deposited polymer layer on the microcantilever resonant frequency and mass sensitivity are presented. Fabrication of the chemical sensor, including a description of polymer deposition through evaporation-driven assembly within a capillary, is detailed. The completely functionalized resonator demonstrates a limit of detection of 1.6 ppm for toluene. An optimal polymer sensitive layer deposition of 42% of the total beam length is measured from frequency instability and sensitivity tests.
No takes yet. Share an insight, caveat, or question.
Dorsey et al. (2014) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: