Key result
A rapid liquid-chromatographic procedure can detect indomethacin in 100-microL neonatal serum samples at concentrations as low as 50 micrograms/L with 100% relative recovery.
Why the study?
Does a rapid HPLC method accurately measure indomethacin levels in neonates with patent ductus arteriosus?
Does a rapid HPLC method accurately measure indomethacin levels in neonates with patent ductus arteriosus?
A rapid and accurate HPLC method for measuring indomethacin in small serum samples from neonates can improve the pharmacological management of patent ductus arteriosus.
May enable neonatal indomethacin monitoring with minimal blood; leaves open clinical impact on patent ductus arteriosus management.
This rapid, accurate "high-performance" liquid-chromatographic procedure is intended for measuring indomethacin in serum from neonates who are receiving indomethacin for symptomatic patent ductus arteriosus. Indomethacin and an internal standard (flufenamic acid) are extracted from serum or plasma with chloroform or diethyl ether at pH 5.0. For the chromatography we used a Waters' Radial Compression Separation System (Radial-NOVA PAK C18 reversed-phase column) and a mobile phase of methanol/sodium acetate buffer (10 mmol/L, pH 3.6), 70/30 by vol. The column effluent is monitored at 254 nm. Both indomethacin and flufenamic acid are eluted within 7 min. Indomethacin can be detected in concentrations as low as 50 micrograms/L, in 100-microL samples. Response varies linearly with indomethacin concentration to at least 2 mg/L. Analytical recovery is 75%; relative recovery is 100%. Precision is excellent. Using this method, we were able to improve the success rate for pharmacological management of symptomatic patent ductus arteriosus, especially in neonates with fast clearance rates for the drug.
No takes yet. Share an insight, caveat, or question.
Ou et al. (1984) studied Patent ductus arteriosus. Liquid-chromatographic determination of indomethacin was evaluated on Detection limit and recovery of indomethacin. A rapid liquid-chromatographic procedure can detect indomethacin in 100-microL neonatal serum samples at concentrations as low as 50 micrograms/L with 100% relative recovery.