A modification of commercially manufactured atomic force microscopy cantilevers which reduces the bending of the V-shaped legs due to changes in temperature is described. Gold-coated silicon nitride cantilevers are a bimorph system in which the different thermal expansion coefficients of the materials comprising the system can produce a temperature-dependent change in curvature. Other stress-related effects might also be responsible for the observed bending. By removing the gold film and redepositing gold only at the end of the V-shaped legs, a reduction in the bending of the cantilever is accomplished while the required optical reflectivity for the laser deflection system is retained. Imaging x-ray photoelectron spectroscopy, scanning electron microscopy, and changes in the detector photodiode signal related to bending of the cantilever are shown for modified and unmodified tips.
No takes yet. Share an insight, caveat, or question.
Wenzler et al. (1996) studied this question.