Data are presented on patient doses measured during standard hospital routine in seven radiological departments in the Province of Brescia. This study is part of a Quality Assurance Programme, carried out to assess the possibility and validity of a regional protocol. Before collecting dose data, tests on the performance of the X ray units and processors were performed in every department according to a Quality Control Protocol. The following examinations were considered: chest, knee, lumbar spine, pelvis, skull and barium meal. The surface entrance doses of 314 adults and 216 children were measured. The sample size for barium meal was lower: 65 adults and only 10 children. The patients anthropometric data and the technical parameters used were collected at the same time. For adults the organ doses and effective dose equivalent (EDE) were calculated. A side range of entrance doses were obtained both for adults and children. The reasons can be: patient size, performance of the equipment and processors, film-screen combination, use of AEC, use of fluoroscopy and grid, training and skill of the staff.