A novel electrochemical aptasensor incorporating a signal enhancement for the determination of cocaine was designed. Gold nanoparticles were self‐assembled onto the surface of a gold electrode through 1,6‐hexanedithiol. A bifunctional derivative of the 32‐base cocaine‐binding aptamer with a redox‐active ferrocene moiety and a thiol linker group at the termini of the strand was self‐assembled onto the surface of gold nanoparticles. The oxidation peak current is linearly related to the concentration of cocaine from 1.0 to 15.0 μM with a detection limit of 0.5 μM. It was found that the sensitivity of the aptasensor with gold nanoparticles modification was ca. 10‐fold higher than that of the aptasensor without gold nanoparticles modification. This work demonstrates that gold nanoparticles‐assembled gold electrode provides a promising platform for immobilizing aptamer and enhancing the sensitivity.
No takes yet. Share an insight, caveat, or question.
Li et al. (2008) studied this question.