Key result
Two new types of bipolar cuff microelectrodes using electro-conductive rubber or water-absorbent polymer successfully recorded renal sympathetic nerve activity in anesthetized rats.
Novel implantable microelectrodes using electro-conductive rubber or water-absorbent polymer allow for effective recording of sympathetic nerve activity in small animals.
Enables refined sympathetic recordings in rat models; leaves open validation for chronic or translational use.
We have developed two new types of bipolar cuff microelectrodes (volume size, 2-3 mm(3)) using electro-conductive rubber or water-absorbent polymer, either of which can be applied to measure sympathetic nerve activity in small animals. A renal nerve bundle of an anesthetized rat was inserted into the center hole of the electrode (diameter, 0.15 mm) through a slit and had good contact with the electrodes. Renal sympathetic nerve activity, which was verified by sympathetic ganglionic blockade, could be recorded using either type of implantable cuff electrode.
No takes yet. Share an insight, caveat, or question.
Matsukawa et al. (2003) studied this question. Bipolar cuff microelectrodes (electro-conductive rubber or water-absorbent polymer) was evaluated on Recording of renal sympathetic nerve activity (RSNA). Two new types of bipolar cuff microelectrodes using electro-conductive rubber or water-absorbent polymer successfully recorded renal sympathetic nerve activity in anesthetized rats.