The design of a controller which compensates for the effects of creep, hysteresis, vibration, and cross-coupling in a piezoelectric tube scanner (PTS) is presented in this paper. The PTS is a key nanopositioning component installed in a commercial atomic force microscope (AFM) to perform scanning. The impediments to fast scanning due to PTS dynamics are: 1) the presence of mechanical resonances; 2) nonlinearities due to the piezoelectric characteristics; and 3) the cross-coupling effect between x- and y-axes in the PTS. In this paper, a multi-input multi-output model predictive control (MPC) scheme is designed to counteract the effects of creep, hysteresis, vibration, and cross-coupling in a PTS. Also, a damping compensator is included to suppress the vibration effect at its resonance frequency. The proposed controller achieves a high closed-loop bandwidth and significant damping of the resonant mode. To evaluate the performance improvement using the proposed control scheme, an experimental comparison with the existing AFM proportional-integral (PI) controller and a single-input single-output (SISO) MPC is conducted. Enhancement in the scanning speed up to 125 Hz is observed with the proposed controller.
No takes yet. Share an insight, caveat, or question.
Raña et al. (2014) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: