Key result
The multi-modal sensor achieved high accuracy for heart beat intervals (cECG RMSE 2.3 ms, rPPG RMSE 18.9 ms; both correlation 0.99) and breathing intervals (MI RMSE 1.12 s, correlation 0.90).
A novel bed-integrated multi-modal sensor demonstrated high accuracy in measuring heart rate and breathing intervals in a small pilot study of healthy subjects.
Answers are grounded in this paper and cite their sources.
May enable unobtrusive vital-sign monitoring; leaves open validation in clinical populations and outcome impact.
In this paper, we present a novel unobtrusive multi-modal sensor for monitoring of physiological parameters featuring capacitive electrocardiogram (cECG), reflective photoplethysmogram (rPPG), and magnetic induction monitoring (MI) in a single sensor. The sensor system comprises sensor nodes designed and optimized for integration into a grid-like array of multiple sensors in a bed and a central controller box for data collection and processing. Hence, it is highly versatile in application and suitable for unobtrusive monitoring of vital signs, both in a professional setting and a home-care environment. The presented hardware design takes both inter-modal interference between cECG and MI into account as well as intra-modal interference due to cross talk between two MI sensors in close vicinity. In a lab study, we evaluated a prototype of our new multi-modal sensor with two sensor nodes on four healthy subjects. The subjects were lying on the sensors and exercising with a hand grip in order to increase heart rate and thus evaluate our sensor both during changing physiological parameters as well as a wider range of those. Heart beat intervals and heart rate variability were derived from both cECG and rPPG. Breathing intervals were derived from the MI sensor. For heart beat intervals, we achieved an RMSE of 2.3 ms and a correlation of 0.99 using cECG. Similarly, using rPPG, an RMSE of 18.9 ms with a correlation of 0.99 was achieved. With regard to breathing intervals derived from MI, we achieved an RMSE of 1.12 s and a correlation of 0.90.
No takes yet. Share an insight, caveat, or question.
Yu et al. (2019) studied Healthy (n=4). Multi-modal sensor (cECG, rPPG, MI) was evaluated on Heart beat intervals and breathing intervals accuracy. The multi-modal sensor achieved high accuracy for heart beat intervals (cECG RMSE 2.3 ms, rPPG RMSE 18.9 ms; both correlation 0.99) and breathing intervals (MI RMSE 1.12 s, correlation 0.90).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: