A brief survey is given of the calibration methods for hydrophones in the ultrasonic frequency range. The methods presently used in the Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, West Germany, for hydrophone calibrations in the frequency range from 1 to 15 MHz are the primary subject of concern. These methods are the two-transducer reciprocity method for the calibration at discrete frequencies, the time-delay-spectrometry substitution method for quasifrequency continuous calibrations, and the two-transducer reciprocity method with time-delay spectrometry, also for quasi-frequency continuous calibration. Compared with the calibration at discrete frequencies, the expenditure of time for a calibration is considerably reduced in the case of the last-mentioned method. The influencing parameters which affect the evaluation of the measurement uncertainty are briefly discussed for the calibration methods applied at the PTB.