An electrochemical method based on the concept of a biosensor for the monitoring of ozone is described for first time. The proposed method includes two parts: a selective sorbent for ozone, that is, eugenol, and a formaldehyde amperometric biosensor mounted into a flow-through cell. Ozone adds rapidly to the double bond of the allyl group of eugenol, which has been immobilized onto a hydrophobic C-18 reactor and the so produced formaldehyde is collected into the working buffer solution (sampler) and pumped to the detector. A multimembrane assembly consisting of an alcohol oxidase-modified nylon membrane sandwiched between an outer polycarbonate and an inner cellulose acetate membrane was fitted onto a Pt electrode and the enzymatically produced H2O2 was monitored at +0.65 V (vs Ag/AgCl/KCl 3 M). Under optimum conditions, a linear calibration curve over the concentration range 3-200 microg x mL(-1) ozone was constructed. The detection limit (S/N = 3) was calculated at 1.1 microg x mL(-1) ozone. The proposed method is interference-free from other gases such as O2, Ar, N2, N2O, NOCl, SO2, NH3, and CO2, which were tested at concentrations >200-fold higher than that of 100 microg x mL(-1) ozone used for comparison. Besides selectivity, the method is easy to perform and reproducible; its applicability in synthetic gaseous samples is also demonstrated.
No takes yet. Share an insight, caveat, or question.
Stergiou et al. (2006) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: