Key result
ST-segment elevation occurred in 4 of 43 patients (9.3%) during pulmonary vein ablation with a circumferential coiled microelectrode catheter, resolving completely without cardiovascular damage.
Why the study?
Does radiofrequency catheter ablation using the 4F REVELATION Helix microcatheter cause ST-segment-elevation in patients with paroxysmal atrial fibrillation?
Does radiofrequency catheter ablation using the 4F REVELATION Helix microcatheter cause ST-segment-elevation in patients with paroxysmal atrial fibrillation?
Transient ST-segment elevation may occur during left atrial and pulmonary vein mapping/ablation for PAF, but appears reversible and without permanent cardiovascular damage.
Answers are grounded in this paper and cite their sources.
Transient ST elevation may occur with this microcatheter but resolves without damage; leaves open comparative safety versus standard catheters.
AIMS: Paroxysmal atrial fibrillation (PAF) is predominantly triggered by focal ectopies located within the pulmonary veins (PV). The BITMAP Study (Breakthrough and Isolation Trial: Mapping and Ablation of Pulmonary Veins) investigated prospectively the safety and efficacy of a catheter design with circumferential mapping and ablation electrodes. We report the phenomenon of ST-segment-elevation during catheter placement in the left atrium (LA) and superior PVs in this multi-centre study. METHODS AND RESULTS: Forty-three patients (57+/-10 years) with PAF were included in this study. Radiofrequency catheter (RFC) ablation supported by the 4F REVELATION Helix microcatheter (Cardima Inc., Freemont, CA, USA) with eight distal-coiled microelectrodes for bipolar mapping and ablation. RFC was applied at the ostial region of PV (30 W, 45-50 degrees C) with a maximum of four RFC applications per electrode. In four of the 43 patients from three centres, we recorded the occurrence of ST-segment-elevation greater than 0.2 mV and accompanying left thoracic discomfort. The ECG changes and the symptoms started abruptly and lasted for 4.2+/-2.2 min. Pericardial effusion could instantaneously be excluded by echocardiography in all cases. Coronary angiograms were performed in three patients with the longest episodes; no thrombotic material or air emboli were present. The symptoms and the ECG changes resolved completely in all patients. CONCLUSION: The phenomenon of ST-segment-elevation during LA- and PV-mapping in patients with PAF may be a common occurrence. In this prospective multi-centre trial, we demonstrated the reversibility of this phenomenon; no cardiovascular or cerebral damage was reported during both the procedure and the follow-up. Although the mechanism is still unclear, vasospasm may contribute to this phenomenon because of autonomic dysregulation.
No takes yet. Share an insight, caveat, or question.
Risius et al. (2006) studied Paroxysmal atrial fibrillation (PAF) (n=43). Radiofrequency catheter ablation using the 4F REVELATION Helix microcatheter was evaluated on Occurrence of ST-segment-elevation greater than 0.2 mV and accompanying left thoracic discomfort. ST-segment elevation occurred in 4 of 43 patients (9.3%) during pulmonary vein ablation with a circumferential coiled microelectrode catheter, resolving completely without cardiovascular damage.