L-glutamate sensor based on glutamate oxidase (GLOx) has been fabricated on N-(3-dimethylaminopropyl)-N'-ethylcarbodiimide hydrochloride (EDC) activated thioglycolic acid (TGA) self-assembled monolayer (SAM) on smart bio-device using mediator-free simple phosphate buffer system (PBS), where TGA-SAM is used as an enzyme immobilization background. The glutamate sensor displayed a low detection limit (DL, 0.089 +/- 0.002 microM), a wide linear dynamic range (LDR, 0.1 microM to10.0 mM), good linearity (R = 0.9931), and higher sensitivity (20.75 +/- 1.0/microM)(-1)cm-2), small sample volume (70.0 microl), as well as good stability and reproducibility. The smart-biodevice system exemplified a simple and efficient approach to the assimilation of GLOx and sensing surface, which can provide analytical access to a large group of enzymes for wide range of bioelectrochemical applications in health care fields.
No takes yet. Share an insight, caveat, or question.
Mohammed M. Rahman (2011) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context:
More from Journal of Biomedical Nanotechnology