Key result
Left ventricular pressure showed a fairly good correlation with the motor current of the centrifugal blood pump in a mock circulatory loop, suggesting cardiac function may be estimated from motor current.
In a mock circulatory loop, motor current analysis of a centrifugal blood pump correlates with left ventricular pressure, suggesting a potential method to estimate cardiac function during LV bypass.
Motor current may estimate LV pressure in mock loops; leaves open translation to clinical LVAD monitoring.
In this study, the effects on varying cardiac function during a left ventricular (LV) bypass from the apex to the descending aorta using a centrifugal blood pump were evaluated by analyzing the left ventricular pressure and the motor current of the centrifugal pump in a mock circulatory loop. Failing heart models (preload 15 mm Hg, afterload 40 mm Hg) and normal heart models (preload 5 mm Hg, afterload 100 mm Hg) were simulated by adjusting the contractility of the latex rubber left ventricle. In Study 1, the bypass flow rate, left ventricular pressure, aortic pressure, and motor current levels were measured in each model as the centrifugal pump rpm were increased from 1,000 to 1,500 to 2,000. In Study 2, the pump rpm were fixed at 1,300, 1,500, and 1,700, and at each rpm, the left ventricular peak pressure was increased from 40 to 140 mm Hg by steps of 20 mm Hg. The same measurements as in Study 1 were performed. In Study 1, the bypass flow rate and mean aortic pressure both increased with the increase in pump rpm while the mean left ventricular pressure decreased. In Study 2, a fairly good correlation between the left ventricular pressure and the motor current of the centrifugal pump was obtained. These results suggest that cardiac function as indicated by left ventricular pressure may be estimated from a motor current analysis of the centrifugal blood pump during left heart bypass.
No takes yet. Share an insight, caveat, or question.
Daiki Kikugawa (2000) studied Simulated failing and normal heart models. Left ventricular bypass using a centrifugal blood pump was evaluated on Correlation between left ventricular pressure and motor current of the centrifugal pump. Left ventricular pressure showed a fairly good correlation with the motor current of the centrifugal blood pump in a mock circulatory loop, suggesting cardiac function may be estimated from motor current.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: