Answers are grounded in this paper and cite their sources.
We developed a novel method to increase the sensitivity of standard enzyme-linked immunosorbent assay (ELISA) using a multiplexed electrokinetic concentration chip. The poly(dimethylsiloxane) (PDMS) molecular concentrator (1) was used to trap and collect charged fluorescent product of target-bound enzyme turnover reaction of ELISA that occurred in a standard 96 well plate. Detection sensitivities of both prostate specific antigen (PSA) and CA 19-9 (a human pancreatic and gastrointestinal cancer marker) ELISAs in serum are enhanced approximately 100 fold with a low CV of <17%. We also integrated this method with an on-chip bead-based ELISA that lends itself toward a fully automated on-chip diagnostic device. Detection sensitivity of microfluidic bead-based CA 19-9 ELISA in serum is enhanced approximately 65 fold compared to the results without the electrokinetic accumulation step. This chip can be directly applied to enhance the readout sensitivity of a wide range of existing ELISA kits at concentrations below the current detection limit.
No takes yet. Share an insight, caveat, or question.
Cheow et al. (2010) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: