Answers are grounded in this paper and cite their sources.
A healthcare chip for checking the health condition at home has been studied based on the analysis of a trace amount of blood collected by a painless needle. A microcapillary as the main component for the blood flow was fabricated by molding its reverse pattern made of a quartz plate on a cheap and disposable poly(ethylene terephthalate) (PET) plate with an area of 2 cm × 2 cm. The painless needle was fabricated by polishing the tip of a stainless steel tube (SUS) with a 100 µm diameter at an angle of 10 degrees using chemical mechanical polishing (CMP) and subsequently finishing it by electropolishing. Coating the 2-methacryloyloxyethyl phosphorylcholine (MPC) polymer on the inner surface of the capillary effectively suppressed the adsorption of corpuscles. A quartz electroosmosis flow pump embedded in the PET plate introduced the blood into a U-shaped capillary through the needle and the blood was centrifugally separated into the corpuscles and plasma on the chip. The plasma was then conveyed to ion-sensitive field-effect transistors (ISFETs) set on the capillary and the Na + and K + ion concentrations were measured with high selectivity. Finally, a series of chip operations from blood collection to measurements of the pH, Na + and K + ion concentrations was demonstrated.
No takes yet. Share an insight, caveat, or question.
Oki et al. (2003) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: