Key result
Constants a and b for conversion of CAVI VS to CAVI0 were estimated to be 0.842 (95% CI 0.836-0.848) and 0.753 (95% CI 0.721-0.786), respectively, allowing direct calculation of CAVI0.
Observational (n=497)
Answers are grounded in this paper and cite their sources.
The study provides equations to calculate a corrected, blood pressure-independent cardio-ankle vascular index (CAVI0) from conventional CAVI measurements, enabling retrospective and prospective use in clinical research.
Objective : Cardio-ankle vascular index (CAVI) as measured using the VaSera device (CAVI VS , Fukuda Denshi), has been proposed as a stiffness index that does not depend on blood pressure. We have recently shown theoretically that CAVI VS still exhibits blood pressure dependence, and proposed the corrected index CAVI 0 . The present study aims to establish a method of calculating C A V I 0 either (i) from VaSera-reported values of cardiac-brachial and brachial-ankle pulse transit times (tb and tba, respectively) and blood pressure, or (ii) directly from CAVI VS . To derive this method, the relationship among CAVI VS and its scale constants a and b , tb, tba, and blood pressure has to be established. Approach : From data of 497 subjects, eight candidate CAVI parameters were defined and calculated, containing all combinations of left or right tb/tba/blood pressure. Candidates were evaluated through correlation with measured left and right CAVI VS . Correlations were compared statistically. Once the correct candidates were determined, two constants ( a and b ) required for converting CAVI VS to CAVI 0 were estimated through linear regression. Main results : Left and right CAVI VS are calculated using left and right tba; however, both left and right CAVI VS are calculated using right brachial blood pressures and right tb. Constants a and b for conversion of CAVI VS to CAVI 0 were estimated to be 0.842 [0.836 0.848] and 0.753 [0.721 0.786] (mean [95%CI]), respectively. Equations to estimate CAVI 0 from CAVI VS , and to directly calculate CAVI 0 from the VaSera output are provided in this paper, as well as in a directly usable spreadsheet supplement. Significance : Our results permit straightforward calculation of C A V I 0 during a study, as well as retrospective estimation of C A V I 0 from CAVI VS in already published studies or where the original transit time values are not available, paving the way for thorough comparison of CAVI 0 to CAVI VS in clinical and research settings. Novelty and significance Cardio-ankle vascular index (CAVI) as measured using the VaSera device (CAVI VS , Fukuda Denshi), has been proposed as a blood pressure-independent arterial stiffness index. We have recently shown theoretically that CAVI VS still exhibits pressure dependence, and proposed a corrected index, CAVI 0 . In the present study, we derived equations to directly obtain CAVI 0 using data from the VaSera device. Our results permit straightforward calculation of C A V I 0 during a study, as well as retrospective estimation of
No takes yet. Share an insight, caveat, or question.
Spronck et al. (2017) reported an observational. Calculation of CAVI0 from CAVI VS was evaluated on Constants a and b for conversion of CAVI VS to CAVI0. Constants a and b for conversion of CAVI VS to CAVI0 were estimated to be 0.842 (95% CI 0.836-0.848) and 0.753 (95% CI 0.721-0.786), respectively, allowing direct calculation of CAVI0.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: